Agilent Technologies EasyEXPERT Guide de l'utilisateur Page 120

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 228
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 119
8-44 Agilent EasyEXPERT Users Guide Vol. 2, Edition 1
Application Library
QSCV Maximum Measurement Value and Accuracy
Figure 8-24 Calculation Example of Measurement Accuracy: MP/HRSMU, ASU+SMU
Conditions:
QSCV Operating Mode: Normal
Measurement Range: 1 A
Output Range: 20 V
Integration Time: 2, 1, 0.5, 0.3 s corresponding to the lines from top to bottom
Equivalent Parallel Resistance of DUT: 10 M
Guard Capacitance of Measurement Path: 200 pF
0
2
4
6
8
10
12
14
0.001 0.01 0.1 1 10
QSCV Measurement Voltage (V)
Reading Accuracy (%)
1.E-14
1.E-13
1.E-12
1.E-11
1.E-10
1.E-09
1.E-08
0.001 0.01 0.1 1 10
QSCV Measurement Voltage (V)
Offset Accuracy (F)
Vue de la page 119
1 2 ... 115 116 117 118 119 120 121 122 123 124 125 ... 227 228

Commentaires sur ces manuels

Pas de commentaire