Agilent Technologies EasyEXPERT Guide de l'utilisateur Page 135

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 228
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 134
Agilent EasyEXPERT Users Guide Vol. 2, Edition 1 9-9
If You Have a Problem
When You Perform Measurement
SMU Oscillates for Negative Resistance Measurements
If the DUT has negative resistance characteristics, SMUs may oscillate. Because the
positive feedback may be configured by the DUT and the SMUs.
To solve this problem:
For voltage controlled negative resistance device
Connect G in parallel with your DUT to cancel negative resistance. To
obtain an output I-V curve, use the following equation.
IY = I - G V
For current controlled negative resistance device
Connect R in series with your DUT to cancel negative resistance. To obtain
an output I-V curve, use the following equation.
VZ = V - R I
If the resistance of the DUT is less than 1 M, you can use a series resistor
built into the SMU.