Error Message
Agilent EasyEXPERT Operation Error
105935 N1258A selector can make an input-output path. Multiple input
channels cannot be specified.
105936 Invalid value (A) was specified for SMU I/V sweep Linear/Log.
Valid option(s): B
105937 Invalid value (A) was specified for C-V sweep Linear/Log.
Valid option(s): B
105938 The start value and stop value of C-V log sweep must be non-zero and
have same polarity.
105939 SMU cannot set the voltage or current exceeds Setting Limit defined in
the SMU Output Setting Limits tab screen on the Configuration
window.
105940 SMU pulse period must be A s longer than measurement delay time +
measurement time. Measurement Delay Time: B s, Measurement
Time:
C s, Pulse Period: D s
105941 Measurement time must be equal to or less than pulse delay time +
pulse width in case of AUTO measurement delay setting.
Channel: A, Pulse Delay Time: B s, Pulse Width: C s, Measurement
Time: D s
105942 SMU pulse period must be A s longer than pulse delay time + pulse
width. Channel: B, Pulse Delay Time: C s, Pulse Width: D s,
Pulse
Period: E s
105943 Total setting current exceeds the capacity of main frame power supply.
Reduce setting current.
105951 HCSMU detected open sense condition. Connect the cables properly
between HCSMU and DUT, then use Kelvin connection.
105952 HCSMU detected overvoltage condition. HCSMU output terminals
may be open. Connect the cables properly between HCSMU and DUT,
then use Kelvin connection.
105953 MCSMU detected overvoltage condition. MCSMU output terminals
may be open. Connect the cables properly between MCSMU and DUT,
then use Kelvin connection.
105960 Module set to the HCSMU field on the Module Selector tab screen
cannot be set to the Master HCSMU or Slave HCSMU field on the
Dual HCSMU Combination tab screen. Empty the HCSMU field and
then set the Dual HCSMU Combination tab screen.
105970 N1265A Ultra High Current Expander / Fixture is not found.