10-54 Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1
Error Message
Agilent FLEX Execution Error
4402 HVSMU over current or guard abuse emergency occurred.
4403 HVSMU guard abuse emergency occurred.
4404 HVSMU HVPS cannot power-off emergency occurred.
4405 HVSMU V ADC lost emergency occurred.
4406 HVSMU I ADC lost emergency occurred.
4407 HVSMU Float lost emergency occurred.
4408 HVSMU HVPS cannot power-on emergency occurred.
5301 Specified module is already assigned to voltage control, current control
or gate control.
Specify a free MCSMU/HCSMU.
5302 Voltage control module and current control module must be different.
Specify a free MCSMU/HCSMU.
5303 Do not execute CN/CNX to the current control module set to
ERPFUHCA.
Execute the CN/CNX command to the module that is set to voltage
control by the ERPFUHCA command.
5304 Ultra High Current Expander / Fixture drain path must be open state.
Open the selector connection path before switch the HVSMU series
resistor.
5305 Execute CL to the module set to ERPFUHCA.
Execute the CL command to the modules that will be specified by the
ERPFUHCA command.
5306 Ultra High Current Expander / Fixture gate path must be open state.
Open the gate connection path before switching the series resistor for
the path.
5307 CRC for N1265A EEPROM Vs correction segment failed.
N1265A might be defective. Contact your nearest Agilent Technologies
service center.
5308 CRC for N1265A EEPROM Is correction segment failed.
N1265A might be defective. Contact your nearest Agilent Technologies
service center.