Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1 8-23
Application Library
QSCV Maximum Measurement Value and Accuracy
Table 8-4 Constant Value for Calculating the Measurement Accuracy: HPSMU
Measurement
Range Constant
Voltage Output Range
QSCV Operating
Mode
2 V 20 V 40 V 100 V 200 V
1 nA Ap 0.11 0.11 0.12 0.12 0.14
Bp 0.033 0.149 0.293 0.717 1.459 Normal
0.0100 0.03 0.05 0.12 0.2347 4155C/4156C
compatible
Cp 0.026
Dp 0.014
Ao 0.03 Normal
0.04 4155C/4156C
compatible
Bo 1.7E-15
Boc 1.3.0E-04
Co 5.9E-14
Coc 4.5.0E-04
Do 1.7E-15
Doc 1.3.0E-04
10 nA Ap 0.11 0.11 0.12 0.12 0.14
Bp 0.033 0.149 0.293 0.717 1.459 Normal
0.0100 0.03 0.05 0.12 0.2347 4155C/4156C
compatible
Cp 0.031
Dp 0.015
Ao 0.03 Normal
0.04 4155C/4156C
compatible
Bo 1.7E-15
Boc 7.3E-05
Co 3.9E-14
Coc 1.9.0E-04
Do 1.7E-15
Doc 7.3E-05