10-48 Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1
Error Message
Agilent FLEX Execution Error
2056 High temperature H/W SRQ detected.
Specified module stopped the output and opened the output relay.
2101 Specified load impedance is out of absolute limits.
Set the appropriate impedance value to SER.
2103 Specified period is out of absolute limits.
Set the appropriate pulse period value to SPPER.
2104 Specified trigger count is out of absolute limits.
Set the appropriate count value to SPRM.
2105 Specified load voltage is out of range.
Set the appropriate voltage to SPV or ALW.
2106 Specified load voltage of added amplitude is out of range.
Set the appropriate voltage to SPV for setting the 3-level pulse output.
2107 Specified voltage is out of absolute limits (can't achieve amplitude).
Set the appropriate voltage to SPV. It must be the voltage which can be
applied under the present load impedance condition.
2108 Specified transition time is out of absolute limits.
Set the appropriate value of leading time or trailing time to SPT.
2111 Leading/0.8 must be within Width value.
Set the appropriate value to SPT. The pulse width value must be more
than leading time/0.8 value.
2112 Delay + Width + Trail/0.8 must be within Period value.
Set the appropriate value to SPT. The pulse period value must be more
than delay time+pulse width+trailing time/0.8 value.
2113 Specified pulse delay is out of absolute limits.
Set the appropriate delay time value to SPT.
2114 Specified pulse width is out of absolute limits.
Set the appropriate pulse width value to SPT.
2115 Specified duration is out of absolute limits.
Set the appropriate duration value to SPRM.