17. Image, Multiple, and Out of Rand Responses Test
9. Repeat steps 4 through 9 for
all
remaining (CENTER FREQUENCY]
and synthesized sweeper settings in
Table
2-20. Steps 4 through 8
need only be done once for each (CENTER FREQUENCY) setting.
10. Set the synthesized sweeper for an output CW frequency of
5700.000 MHz.
11. Key in the following analyzer settings:
(CENTER
FREQUENCY)
...................................... 5.7 GHZ
(REFERENCE LEVEL]
........................................
0.0
dBm
12. Press MARKER
CPRESEL
PEAK
)
and wait for PEAKING! message to
disappear from display. Press MARKER
(jj].
13. Change
FREQUENCY
SPAN] to 5
kHz.
Adjust the synthesized
sweeper output power level to place peak of signal trace at the
top CRT graticule line.
14. Press MARKER
a].
Using the DATA knob, determine
the
amplitude of the spurious response and enter the result in
Iable
2-21.
15. Change (CENTER FREQUENCY) to 2.36790
GHz.
The multiple
response at the center frequency should be below the
m),
SWEEP
(CONT)’
display line
(s-70
dBc).
Multiple response at 2.36790
GHz
dBc
16. Press
ISHIFT),
(MKR/A+
STP
SIZE)Q.
17. Set the synthesized sweeper output CW frequency and analyzer
(CENTER FREQUENCY) according to Table 2-21. Before checking the
amplitude of the multiples for a given signal frequency, set the
input signal amplitude by the procedure of steps 11 through 14.
‘Ibble
2-2 1. Multiple Responses
I
5700.000
I
2.68930
-70
I
6000.000
I
1.18930
-50
12000.000
8.107133
8.535667
13000.000
1.06790
1.9107
0.53395
15000.000
10.107133
-60
10.535667
-60
Displayed S purious
Amplit
tde
Measured
Maximum
ww
WV
-70
-70
-45
-45
-45
2-76 Performance Tests