Agilent Technologies HP 8566B Spécifications Page 375

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 405
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 374
5.
Impulse
and
Resolution
Bandwidth
Switching
Uncertainty
Tkst
Related
Adjustment
Specification
Description
Equipment
Procedure
3 MHz Bandwidth Filter Adjustments
21.4 MHz Bandwidth Filter Adjustments
Down/Up Converter Adjustments
f2.0
dB,
10 Hz bandwidth
*0.8
dB,
30 Hz bandwidth
f0.5
dB,
100 Hz to 1 MHz bandwidth
fl.O
dB,
3 MHz bandwidth
30
kHz
and 100
kHz
bandwidth switching uncertainty figures only
applicable
590%
Relative Humidity.
The CAL OUTPUT signal is applied to the input of the spectrum
analyzer. The deviation in peak amplitude of the signal trace is then
measured as each resolution bandwidth filter is switched in.
None required
1.
Press
[2--22].
2. Connect CAL OUTPUT to RF INPUT.
3. Key in the following control settings:
[CENTER FREQUENCY)
.......................................
100
[FREQUENCY SPAN)
...........................................
.5
REFERENCE LEVEL)
&ii,
......................................... .-8
.....................................................
1
4. Press LOG [ENTER
dB/olv)
and key in 1
dB.
Press MARKER
(PEAK SEARCH)
a.
MHz
MHz
dBm
MHz
5. Key in settings according to
Table
4-6. Press MARKER
[PEAK SEARCH) at each setting, then read the amplitude deviation
from the MARKER La] readout at the upper right of the display
(see Figure
4-6).The
allowable deviation for each resolution
bandwidth setting is shown in the table.
4-16
Option 462
Vue de la page 374
1 2 ... 370 371 372 373 374 375 376 377 378 379 380 ... 404 405

Commentaires sur ces manuels