Agilent Technologies HP 8566B Spécifications Page 363

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 405
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 362
3.
Impulse
and
Resolution
Bandwidth
Accuracy
T&t
Related
Adjustment
Specification
Description
SYNTHES I ZEA
LEVEL GENERATOR
Impulse Bandwidth Adjustments
f20%,
3 MHz bandwidth
f
1096,
1 MHz to 1
kHz
bandwidths
-0,
+50%,
300 Hz to 10 Hz (6
dB
bandwidths)
A frequency synthesizer and pulse/function generator are used
to input pulses to the spectrum analyzer. The amplitude of the
pulses is measured, and the impulse bandwidths are calculated for
each impulse bandwidth from 3 MHz to 1
kHz.
The 6
dB
resolution
bandwidths are then measured using the spectrum analyzer
(JiiKK]
function. The CAL OUTPUT signal is used as a stable signal source to
measure the 6
dB
resolution bandwidths.
PULSE/FUNCTION
GENERATOR
I
EXT
I
OUTPUT
INPUT
SPECTRUM ANALYZER
Figure 4-2. Impulse Bandwidth Test Setup
Equipment
Frequency Synthesizer
.....................................
HP 3335A
Pulse/Function Generator
..................................
HP 8116A
4.4 Option 462