2 1. Frequency Response Adjustments
Note
SWEEP TIME
.............................................
.30
ms
SWEEP
....................................................
CONT
148. On the spectrum analyzer, key in TRACE A [CLEAR-WRITE),
(SWEEP
TIME)
2s,
MARKER (OFF),
I=].
As the spectrum
analyzer completes each sweep, a series of approximately 23 new
responses should be displayed. The peaks of these responses
coarsely outline the spectrum analyzer frequency response.
149. Gradually adjust (Band E Lower Segment) A6AllR60 El and
(Band E Upper Segment)
A6AllR78
E2 for maximum flatness of
the displayed signal responses. The adjustments are interactive,
with A6AllR60 El having the most effect on the level of the
displayed signals below mid-band (approximately 20.3
GHz),
and A6AllR78 E2 having the most effect on the level of
the displayed signals above mid-band. Adjust A6AllR60 El
counterclockwise to increase the level of the displayed signal
responses below mid-band. Adjust
A6AllR78
E2 clockwise
to increase the level of the displayed signal responses above
mid-band. See Figure 3-90 for the locations of A6AllR60 El and
A6Al
lR78
E2.
If the displayed signals at the high end of the band drop off by more
than 3-4
dB,
center A6AllR78 E2 and proceed to step 150.
150. On the spectrum analyzer, key in [SWEEP TIME)
5s,
LOG SCALE
C
ENTER
dB/DIv]
1
dB,
TRACE B (VIEW),
IsAvE_)
4,
C-1
2s,
TRACE B
[BLANK),
(HOLD].
151. Readjust 17-turn (YTX Diode Bias) potentiometer
A6AlOR95
VE
to maximize the overall level of the displayed signal responses
from 18.6
GHz
to 22.0
GHz
on the spectrum analyzer display.
Then, adjust
A6AlOR95
VE clockwise until signal responses at
the high end of the band drop in amplitude by approximately 1.5
dB.
152. Readjust 22-turn (IF Offset) potentiometer
A6A12R82
E as
necessary to maximize the
overall
level of the displayed
signal
responses from 18.6
GHz
to 22.0
GHz
on the spectrum analyzer
display. If the displayed
signal
responses peak at widely different
settings of
A6A12R82
E, perform the following steps:
a. Adjust 22-turn (YTX Linearity) potentiometers
A6AlOR70
LB4
and
A6AlOR76
LR4 fully counterclockwise if they have not
already been so adjusted; do not readjust
A6A12R40
LBl,
A6AlOR41
LB2,
A6AlOR42
LB3,
A6AlOR31
LRl,
A6A610R34
LR2, or
A6AlOR37
LR3.
b. Readjust 22-turn (IF Offset) potentiometer
A6A12R82
E as
necessary to maximize the overall level of the displayed signal
responses at the low end of the band.
c. On the spectrum analyzer, key in MARKER
(NORMAL)
and
position the marker at the point in the band where the signal
responses begin to fall off from their maximum value.
d. Readjust 22-turn (YTX Linearity) potentiometer
A6AlOR70
LB4 clockwise until the displayed signal responses at the high
end of the band begin to drop. Continue to adjust
A6AlOR70
LB4 until the
rolloff
point aligns with the position of the
Adjustments
3-l