119.
On the spectrum analyzer, key in LOG SCALE (ENTER
dB/Div)
2
dB,
MARKER
L@Z%Zj.
120.
Readjust 22-turn (IF Offset) potentiometer
A6A12R83
D as
necessary to maximize the amplitude of the 15.0
GHz
signal on
the spectrum analyzer display.
121.
Set the synthesized sweeper controls as follows:
START FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.12.5
GHz
STOP FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18.6
GHz
RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
LEVELING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . XTAL
SWEEP TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.30
ms
SWEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CONT
122.
On the spectrum analyzer, key in TRACE A [CLEAR-WRITE),
(
SWEEP TIME
)
2s,
MARKER (OFF), (HOLD. As the spectrum
analyzer completes each sweep, a series of approximately 22 new
responses should be displayed, as shown in Figure 3-98. The
peaks of these responses coarsely outline the spectrum analyzer
frequency response.
123.
Gradually adjust (Band D Lower Segment) A6AllR57 Dl and
(Band D Upper Segment)
A6AllR75
D2 for maximum flatness of
the displayed signal responses. The adjustments are interactive,
with A6AllR57 Dl having the most effect on the level of
the displayed signals below mid-band (approximately 15.5
GHz),
and A6AllR75 D2 having the most effect on the level of
the displayed signals above mid-band. Adjust A6AllR57
Dl
counterclockwise to increase the level of the displayed signal
responses below mid-band. Adjust
A6AllR75
D2 clockwise
to increase the level of the displayed signal responses above
mid-band. See Figure 3-90 for the locations of A6AllR57
Dl
and
A6Al
lR75
D2.
2 1. Frequency Response Adjustments
the displayed 15.0
GHz
signal. Adjust 17-turn (Band D Step
Gain) potentiometer
A6AlOR27
GD as necessary to adjust the
amplitude of the displayed 15.0
GHz
marker to -10.00
dBm
fO.10
dB.
Adjust
A6AlOR27
GD counterclockwise to increase
the signal level, and clockwise to decrease the signal level.
If
A6AlOR27
GD does not have sufficient range, adjust the
amplitude of the displayed 15.0
GHz
marker as close as possible
to -10.00
dBm.
See Figure 3-90 for the location of
A6AlOR27
GD.
Note
If the displayed signals at the high end of the band drop off by more
than 2-3
dB,
center A6AllR75 D2 and proceed to step 125.
124. On the spectrum analyzer, key in
@WEEP
TIME)
5s,
LOG SCALE
[
ENTER
dB/Div)
1
dB,
TRACE B [VIEW), (SAVE) 4,
c-1
2s,
TRACE B [BLANK),
[ml.
125. Readjust 17-turn (YTX Diode Bias) potentiometer
A6AlOR12
VD
to maximize the overall level of the displayed signal responses
from 12.5
GHz
to 18.6
GHz
on the spectrum analyzer display.
Then, adjust
A6AlOR12
VD clockwise until signal responses at
Adjustments 3-195