12. 10 MHz Standard Adjustment (SN 2637A and Below)
message should typically appear on the spectrum analyzer display for
10 minutes or less after line power is first applied to the RF Section.
Note
The rear-panel FREQ REFERENCE switch enables or disables the RF
Section +20 Vdc switched supply, which powers the oscillator circuits
in the A22 10 MHz Frequency Standard. This switch must be set to
INT and the spectrum analyzer must be switched ON continuously
(not in STANDBY) for at least 72 hours before adjusting the frequency
of the A22 10 MHz Frequency Standard.
2. Set the LINE switch to ON. Leave the spectrum analyzer ON
(not in STANDBY) and undisturbed for at least 48 hours to allow
the temperature and frequency of the A22 10 MHz Frequency
Standard to stabilize.
Note
‘lb prolong CRT life, press
ISHIFT) [CLEAR-WRITES
to turn off the CRT
display while the spectrum analyzer is unattended, and
(m’
[my
to turn the CRT back on.
3. Connect the (Cesium Beam) Frequency Standard to the Frequency
Counter rear-panel
TIMEBASE
IN/OUT connector as shown in
Figure 3-52.
T
IMBASE
FREQUENCY
COUNTER
IN/OUT
FREQUENCY
STANDARD
l9
Figure 3-52. 10 MHz Frequency Standard Adjustments Setup
4. Disconnect the short jumper cable on the RF Section rear panel
from the FREQ REFERENCE INT connector. Connect this output
(FREQ REFERENCE INT) to INPUT A on the Frequency Counter.
A REF UNLOCK message should appear on the CRT display.
5. Set the Frequency Counter controls as follows:
FUNCTION/DATA
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . FREQ A
INPUT A:
x10ATTN
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
..OFF
AC
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OFF (DC coupled)
500
Z . . . . . . . . . . . . . . . . . .
.OFF
(1
MO
input impedance)
AUTO TRIG
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ON
100
kHz
FILTER A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OFF
INT/EXT switch (rear panel) . . . . . . . . . . . . . . . . . . . . EXT
6. On the Frequency Counter, select a 10 second gate time by
pressing
C-TIME)
10 [GATE).Offset the displayed
frequency by -10.0 MHz by pressing MATH (SELECT/ENTER
(CHS/EEX)
10
C-1
6 @ELECT/ENTER) (SELECT/ENTER).
T!te
Adjustments
3.105