Frequency Response Adjustments
On the spectrum analyzer, key in
[=I
4,
m
IGHz)
/ and
use the DATA knob to gradually change the PRESELECTOR DAC
setting from 32, maximizing the level of the TRACE A displayed
signal responses at the lowest point on the TRACE B waveform.
Note the PRESELECTOR DAC setting.
On the spectrum analyzer, key in
L-1
6,
LSHIFT)
(GHz)
/
and enter in the PRESELECTOR DAC setting from step 81.
Press SWEEP (SINGLE) and wait for the sweep to complete (150
seconds) and the SWEEP LED to turn off.
Repeat steps 81 and 82 until the level of the lowest point on
the TRACE B waveform does not change. Repeat step 82 with a
PRESELECTOR DAC value of 30 and 34.
On the spectrum analyzer, key in TRACE B [VIEW), MARKER
(NORMAL_)
and use the DATA knob to position a marker on the
lowest point on the TRACE B waveform. Then, press MARKER
Ln],
MARKER (PEAK SEARCH) to position a second marker on
the highest point on the TRACE B waveform. Note the total
peak-to-peak deviation of the displayed trace.
Band B total peak-to-peak deviation:
dB
On the spectrum analyzer, press MARKER
@ZZiXj
and note the
frequency of the highest point on the TRACE B waveform.
Band B highest point:
GHz
Repeat steps 79 through 85 if necessary until the total
peak-to-peak deviation of the TRACE B waveform is less than
3.40
dB.
See the TRACE B reference waveform to indicate which
portions of the frequency response waveform require relative
adjustment.
On the spectrum analyzer, key in
(RECALL)
4,
IHOLD).
On the synthesized sweeper, press
[cw)
and enter the frequency
recorded in step 85, positioning the displayed TRACE A signal
response at the highest point on the TRACE B waveform.
Note
The RF Section front-panel AMPTD CAL control should still be set to
the approximate center of its adjustment range from step 26.
Note
89. On the spectrum analyzer, key in MARKER
(j-1,
MARKER
[PEAK SEARCH) to position a marker at the peak of the displayed
TRACE A signal response. Adjust 17-turn (Band B Step Gain)
potentiometer
A6AlOR23
GB to adjust the amplitude of the
marker to -10.00
dBm
+ (l/2 of Band B total peak-to-peak
deviation)
f0.01
dB.
Adjust
A6AlOR23
GB counterclockwise to
increase the signal level, and clockwise to decrease the signal
level.
If the
A6AlO
Miscellaneous Bias/Relay Driver Assembly is HP P/N
85660-60322 (HP
85660A/B
RF Sections with serial number prefix
2747A or above), and
A6AlOR23
GB does not have sufficient range,
change the value of factory-select component
A6AlOR87.
Increase
the value of
A6AlOR87
to decrease the signal level, and decrease the
value of
A6AlOR87
to increase the signal level. See
‘Iable
3-3 for the
Adjustments 3-189