2 1. Frequency Response Adjustments
acceptable range of values for
A6AlOR87,
and Table 3-4 for HP part
numbers. See Figure 3-90 for the location of
A6AlOR87.
Band
C,
5.8
GHz
to
90. On the spectrum analyzer, key in
@=TEZT@,
@iYiE]
[PRESEL
PEAK_)
12.5
GHz
=,
[SHIFT]
TRACE A
(HOLD]
b.
Set the spectrum analyzer
controls as follows:
START FREQ . . . . . . , . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.5.8
GHz
STOP FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.5
GHz
RES BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.3
MHz
REFERENCE LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . -7
dBm
LOG SCALE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 dB/DIV
SWEEP TIME
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500 ms
91. On the synthesized sweeper, key in
Icw]
6.0
GHz.
Note
The POWER LEVEL of the synthesized sweeper should still be leveled
to a calibrated -10.00
dBm
at 100 MHz at the spectrum analyzer RF
INPUT from step 55.
92.
Adjust 22-turn (IF Offset) potentiometer
A6A12R84
C and
17-turn (YTX Bias) potentiometer
A6AlOR15
VC for maximum
signal amplitude at 6.0
GHz
on the spectrum analyzer display.
See Figure 3-90 for the locations of
A6A12R84
C and
A6AlOR15
vc.
93.
On the spectrum analyzer, key in MARKER
(j-1,
MARKER
[PEAK SEARCH) to position a marker at the peak of the displayed
6.0
GHz
signal. Adjust 17-turn (Band C Step Gain) potentiometer
A6AlOR25
GC as necessary to adjust the amplitude of the
displayed 6.0
GHz
marker to -10.00
dBm
fO.10
dB.
Adjust
A6AlOR25
GC counterclockwise to increase the signal level, and
clockwise to decrease the signal level. If
A6AlOR25
GC does not
have sufficient range, adjust the amplitude of the displayed 6.0
GHz
marker as close as possible to -10.00
dBm.
See Figure 3-90
for the location of
A6AlOR25
GC.
94.
On the spectrum analyzer, key in LOG SCALE (ENTER
dB/DIv]
2
dB,
MARKER
fjNORMAL).
95.
Readjust 22-turn (IF Offset) potentiometer
A6A12R84
C as
necessary to maximize the amplitude of the 6.0
GHz
signal on
the spectrum analyzer display.
96.
Set the synthesized sweeper controls as follows:
START FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.5.8
GHz
STOP FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.5
GHz
RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
LEVELING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XTAL
SWEEP TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30 ms
SWEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CONT
97.
On the spectrum analyzer, key in TRACE A
CCLEAR-WRITE),
(SWEEP TIME)
2s,
MARKER
(OFF,
IHOLD).
As the spectrum
analyzer completes each sweep, a series of approximately 23 new
resnonses
should be disnlaved.
as shown in Figure 3-97. The
3-190
Adjustments
Commentaires sur ces manuels