2 1. Frequency Response Adjustments
Note
If
A6AlO
Miscellaneous Bias/Relay Driver Assembly is HP P/N
85660-60322 (HP 85660AB RF Sections with serial number prefix
2747A or above), and
A6AlOR27
GD does not have sufficient range,
change the value of factory-select component
A6AlOR88.
Increase
the value of
A6AlOR89
to decrease the signal level, and decrease the
value of
A6AlOR89
to increase the signal level. See
Table
3-3 for the
acceptable range of values for
A6AlOR89,
and
Table
3-4 for HP part
numbers. See Figure 3-90 for the location of
A6AlOR89.
Band
E,
18.6
GHz
to
141. On the spectrum analyzer, key in
C-J,
LSHIFT)
[PRESEL
PEAK]
22
GHz
=
,
(-1
TRACE A
(MAX]
b.
Set the spectrum analyzer
controls as follows:
START FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18.6
GHz
STOP FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.22
GHz
RES BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.3
MHz
REFERENCE LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -7 dBm
LOG SCALE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 dB/DIV
SWEEP TIME
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500 ms
142. On the synthesized sweeper, key in
Icw
20.0
GHz.
Note
The POWER LEVEL of the synthesized sweeper should still be leveled
to a calibrated -10.00
dBm
at 100 MHz at the spectrum analyzer RF
INPUT from step 55.
143.
Adjust 22-turn (IF Offset) potentiometer
A6A12R82
E and
17-turn (YTX Bias) potentiometer
A6AlOR9
VE for maximum
signal amplitude at 20.0
GHz
on the spectrum analyzer display.
See Figure 3-90 for the locations of
A6A12R82
E and
A6AlOR95
VE.
144.
On the spectrum analyzer, key in MARKER
@?ZiZiJ,
MARKER [PEAK SEARCH] to position a marker at the peak of
the displayed 20.0
GHz
signal. Adjust Ill-turn (Band E Step
Gain) potentiometer
A6AlOR29
GE as necessary to adjust the
amplitude of the displayed 20.0
GHz
marker to -10.00
dBm
fO.10
dB.
Adjust
A6AlOR29
GE counterclockwise to increase
the signal level, and clockwise to decrease the signal level.
If
A6AlOR29
GE does not have sufficient range, adjust the
amplitude of the displayed 20.0
GHz
marker as close as possible
to -10.00
dBm.
See Figure 3-90 for the location of
A6AlOR29
GE.
145.
On the spectrum analyzer, key in LOG SCALE
C
ENTER
dB/Dw]
2
dB, MARKER (NORMAL).
146.
Readjust 22-turn (IF Offset) potentiometer
A6A12R82
E as
necessary to maximize the amplitude of the 20.0
GHz
signal on
the spectrum analyzer display.
147.
Set the synthesized sweeper controls as follows:
START FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.18.6
GHz
STOP FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.22.0
GHz
RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
LEVELING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XTAL
3-199 Adjustments