10. Frequency Response Test
Procedure
100Hzto
1OOkHz
1. Connect CAL OUTPUT to the RF INPUT on the spectrum analyzer.
2. Press
C-J,
(RECALL) @, MARKER
(
PEAK SEARCH
],
and adjust
AMPTD CAL for a MARKER amplitude of -10.00
dBm
f0.02
dB.
3. Press
(-1
on the spectrum analyzer. Connect function
generator to analyzer RF INPUT as shown in Figure 2-14.
4. Key in analyzer settings as follows:
@TART
FREQ)
................................................
.
kHz
(sTop-
.............................................
..10 0
kHz
5. Set function generator controls as follows:
RANGE Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 K
FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.2,
OFFSET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CAL
AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 V
AMPLITUDE VERNIER . . . . . . . . . . . . . . . . . . . . . . . . . midrange
SYM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CAL
TRIGGER PHASE
. . . . . . . . . . . . . . . . . . . . . . . . . . . FREE RUN
MODULATION
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . all out
MODULATION RANGE Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
MODULATION RANGE Hz VERNIER . . . . . . . . 10 o’clock
MODULATION SYM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CAL
Percent Modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..fullycw
6. Adjust function generator FREQUENCY to place generator signal
near the center graticule on the analyzer display.
7. Adjust the AMPLITUDE VERNIER on the function generator until
the peak of the generator signal is at the reference level line on
the analyzer display.
8. Press LOG [ENTER
dB/plv]
on the analyzer and key in 1
dB
per
division.
9. Adjust function generator AMPLITUDE VERNIER to place peak of
generator signal 2
dB
(2 divisions) down from the reference level.
Do not readjust AMPLITUDE VERNIER during test.
10. Adjust FREQUENCY on the function generator to position the
signal trace at the right edge of the analyzer display.,
11. Press TRACE A
(MAX].
Press MODULATION SWP on
the function generator. When function generator completes
one sweep, press TRACE A (VIEW). Trace should appear as in
Figure 2-15.
2-34
Performance Tests