Agilent Technologies HP 8566B Spécifications Page 308

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 405
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 307
19. CAL Output Adjustment
CENTER
550
MHZ
SPAN
1
an
GHZ
RES
BW
3
MHZ
“BW
3
MHZ
SWP
25.a
msec
Figure 3-86. CAL OUTPUT Harmonics
10. On the second spectrum analyzer, press MARKER [PEAK SEARCH),
[al to position a marker on the peak of the displayed 100
MHz signal. Then, press the
Q)
key (or turn the DATA knob)
to position the second marker on the peak of each displayed
harmonic of the 100 MHz signal in succession, and verify that the
level of each harmonic is greater than 25
dB
down relative to the
peak of the 100 MHz CAL OUTPUT signal. See Figure 3-86.
If the level of each harmonic of the 100 MHz CAL OUTPUT is not
greater than 25
dB
down perform the following steps:
a.
b.
C.
d.
e.
f.
On the second spectrum analyzer, press the
Qf)
and
(7D
keys to
place the second marker on the peak of the highest harmonic
of the 100 MHz signal.
Adjust A6A9AlC29 TRIPLER MATCH only as much as required
to lower the level of the highest harmonic to greater than 25
dB
down relative to the peak of the 100 MHz CAL OUTPUT
signal. Refer to Figure 3-85 for the location of
A6A9AlC29
TRIPLER MATCH.
Disconnect cable 2 (red) from A6A9Jl 300 MHz OUTPUT on
the
A6A9
Phase Lock Assembly. Connect the RF INPUT of the
second spectrum analyzer to
A6A9Jl
300 MHz OUTPUT.
On the second spectrum analyzer, key in
PREFERENCE
LEVEL]
+25
dBm,
[ATTEN]
40
dB,
MARKER
IOFF),
MARKER [PEAK SEARCH) to
position a marker on the peak of the displayed 300 MHz signal.
Readjust A6A9AlC29 TRIPLER MATCH as necessary for a 300
MHz signal level of at least + 16.5
dBm.
Disconnect the second spectrum analyzer from A6A9Jl 300
MHZ OUTPUT, and reconnect cable 2 (red) to
A6A9Jl
300 MHz
OUTPUT.
Repeat steps 9 through
10E
until no further adjustment is
required.
11. Replace the RF Section bottom cover.