17. Image, Multiple, and Out of Rand Responses Test
17.
Image,
Multiple,
and
Out
of
Band
Responses
Tkst
Description
Image and out-of-band responses are checked by setting the analyzer
center frequency to several frequencies across the analyzer range and
tuning a leveled signal source to the frequencies determined by the
tuning equation,
Fsiz
= nFLof F
IF.
Input signals at these frequencies
will excite all possible image and out-of-band responses for a given
1st LO frequency and all positive integer values of n. In this test,
only values of n corresponding to the analyzer mixing modes are
used. Multiple responses are checked by applying an input signal
and measuring the response at those center frequencies for which a
harmonic of the 1st LO mixes with the input signal.
SPECTRUM ANALYZER
Q
ADAPTER
MA
CMLE
ACNIIBLV
lPo1
ADAPl
BNC
‘TEE
‘ER
Figure 2-31. Image, Multiple, and Out-of-Rand Responses Test Setup
Equipment
Synthesized Sweeper
.......................................
HP 8340A
61 cm (24 in.) Cable Assembly, SMA Male Connectors
....
5061-1086
Adapter, Type N (m) to SMA (f)
............................
1250-1250
Adapter, SMA Female Connectors
.........................
.1250-1158
BNC Tee
....................................................
1250-0781
Performance Tests 2-73