15. Residual Responses Test
Figure 2-27. Residual Responses Measurement
8. Look for any residual responses at or above the display line. If a
residual is suspected, press SWEEP
C-j
again and see if the
response persists. A residual will persist on repeated sweeps, but
a noise peak will not. Any residual responses must be below the
display line.
Largest Residual Level
HZ
dBm
at
9. If a response appears to be marginal, perform the following check
to
a.
b.
determine whether or not it exceeds the specification:
Press
ISAVE_)
0.
Press MARKER
(-1
and place the marker on the peak of
the response in question.
C.
d.
Press MARKER
(
SIGNAL
TRACK)
then activate SWEEP
ICONT).
Reduce (FREQUENCY SPAN) to 1 MHz. Reduce
(RESJ
until
there is at least a
‘7
dB
margin between the display line and the
average noise level. The amplitude of the response must be
less than the display line setting.
e.
Press
B
@ to resume the search for residuals.
10. Key in control settings as follows:
START FREQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
GHz
[&FTEj-)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.5GHz
2-62 Performance Tests