7. IF Gain Uncertainty Test
23. Find the largest positive deviation and the largest negative
deviation for reference level settings from 0
dBm
to -70
dBm
in
‘fable
2-9. Also, find the largest positive and negative deviations
for the last five settings in the table.
I
A
Reference Level Range:
0 to -70
dBm
Largest Positive Deviation:
dB
Largest Negative Deviation:
dB
B
-80 to -120
dBm
dB
dB
24. Find the largest positive and negative deviations in Table 2-10 and
Table
2-11:
Largest Positive Deviation:
Largest Negative Deviation:
C
‘able
10
D
Yhble
11
dB
dB
dB
dB
25. The sum of the positive deviations recorded in A, C, and D should
not exceed 0.6
dB.
26. The sum of the negative deviations recorded in A, C, and D
should not be less than -0.6
dB.
27. The sum of the positive deviations recorded in A, B, C, and D
should not exceed 1.0
dB.
28. The sum of the negative deviations recorded in A, B, C, and D
should not exceed
-
1 .O
dB.
2-26 Performance Tests