11. Down/Up Converter Adjustments
Note
The adjustment of A4A6AlR29 WIDE GAIN is affected by
factory-select component
A4A7R60,
which sets the overall gain of
the
A4A7
3 MHz Bandwidth Filter Assembly. If the
A4A7
3 MHz
Bandwidth Filter Assembly or the
A4A5
Step Gain Assembly is
repaired or replaced, perform Adjustment Procedure 10, “Step Gain
and 18.4 MHz Local Oscillator Adjustments” to select (if necessary)
A4A7R60
before adjusting A4A6AlR29 WIDE GAIN (steps 14-17).
1. Position the spectrum analyzer upright as shown in Figure 3-50
and remove the IF-Display Section top cover.
2. Set the spectrum analyzer LINE switch to ON and press
[GHz_).
3. Connect the spectrum analyzer CAL OUTPUT to the RF INPUT.
4. On the spectrum analyzer, key in [RECALL-] 8 and adjust the
front-panel AMPTD CAL control for a displayed signal level of
-10.00
dBm.
5. On the spectrum analyzer, key in
(ml
9 and adjust the
front-panel FREQ ZERO control for maximum amplitude of the
displayed signal.
A4A6
DOWN/UP CONVERTER
A4A6
Figure 3-51. Location of Down/Up Converter Adjustments
Down
Converter
Gain
Adjustment
Note
Perform steps 6 through 15 if the
A4A6
Down/Up Converter Assembly
has been repaired, or if factory-select component
A4A7R60
has
insufficient adjustment range. Otherwise, skip to step 17.
6. Set the spectrum analyzer LINE switch to STANDBY. Remove
A4A6
Down/Up Converter Assembly from the IF-Display Section
and install on two extender boards.
7. Set the spectrum analyzer LINE switch to ON, and key in
(2-22GHz),
(CENTER FREQUENCY) 100 MHZ,
(FREQUENCY
SPAN] 0 HZ,
(mBW)
1
kHz,
[REFERENCE LEVEL) -10
dBm.
8. Connect the active probe to the probe power supply and the
RF INPUT of the second spectrum analyzer, as shown in Figure
3-50. On the second spectrum analyzer, key in
1-1,
Adjustments
3-l