10.
Step
Gain
and
18.4
MHz
Local
Oscillator
Adjustments
Reference
Related
Performance
Tests
Description
IF-Display Section
A4A7
3 MHz Bandwidth Filter
A4A5
Step Gain
Resolution Bandwidth Selectivity Test
IF Gain Uncertainty Test
Center Frequency Readout Accuracy Test
First, the IF signal from the RF Section is measured with a power
meter and adjusted for proper level. Next, the 10
dB
gain steps are
adjusted by connecting the CAL OUTPUT signal through two step
attenuators to the RF INPUT and keying in the REFERENCE LEVEL
necessary to activate each of the gain steps, while compensating for
the increased gain with the step attenuators. The 1
dB
gain steps
are checked in the same fashion as the 10
dB
gain steps, and then
the variable gain is adjusted. The 18.4 MHz oscillator frequency is
adjusted to provide adequate adjustment range of front-panel FREQ
ZERO control; and last, the +
1OV
temperature compensation supply
used by the
A4A4
Bandwidth Filter and
A4A8
Attenuator-Bandwidth
Filter is checked and adjusted if necessary.
ATTENUATOR
ATTENUATOR
DIGITAL
VOLTMETER
Figure 3-46.
Step Gain and 18.4 MHz Local Oscillator Adjustments Setup
3-94
Adjustments