Agilent Technologies HP 8566B Spécifications Page 298

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 405
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 297
17. 20/30 Loop Phase Lock Adjustments
12. Set the controls of the second spectrum analyzer as follows:
CENTER FREQUENCY
..................................
166 MHz
FREQUENCY SPAN
......................................
.50
MHz
REFERENCE LEVEL
................................... -14
dBm
MARKER
................................................
(j-j
Note
Adjustment AlOA4Lll
PLL3
VCO ADJ is sealed at the factory and
normally does not require readjustment unless a component failure
has occurred.
AlOA4Lll
PLL3 VCO ADJ should be adjusted with
the metal shield installed over the PLL3 VCO on
AlOA4
PLL3
Up
Converter Assembly.
13. A 166 MHz signal should be centered on the display of the second
spectrum analyzer, indicating that PLL3 is phaselocked. If not,
coarse adjust AlOA4Lll
PLL3
VCO ADJ to center the signal
displayed on the second spectrum analyzer and phaselock PLL3.
14. With PLL3 phaselocked, readjust
AlOA4Lll
PLL3 VCO ADJ as
necessary for a DVM indication of -7.0
f
0.1 Vdc.
15. On the RF Section, key in
CCENTER
FREQUENCY] 42.569999 MHz,
[
FREQUENCY SPAN
]
0
HZ
,
S
WEEP
(SINGLE),
TR
AC
E A
(CLEARWRITE].
The DVM indication should be -3.5
f
0.6 Vdc.
16. On the RF Section, key in (CENTER FREQUENCY) 42.6499
MHz, (FREQUENCY SPAN) 100
kHz,
SWEEP
(?iIGZ],
TRACE A
(CLEARWRITE).
17. Set the controls of the second spectrum analyzer as follows:
CENTER FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.164.001
MHz
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 MHz
REFERENCE LEVEL
. . . . . . . . . . . . . . . . . . . -14
dBm
18. On the second spectrum analyzer, press MARKER [PEAK SEARCH)
and verify that the output power level of the PLL3 VCO at
164.001 MHz is -20
dBm
f
2
dB
as indicated by the marker.
If it is not, note the amplitude and change the values of
factory-select resistors
AlOA4R29
and
AlOA4R33
as necessary.
Refer to
‘Iable
3-13 for a list of standard resistor values and
corresponding change in circuit gain to adjust the PLL3 VCO
output power level to -20
dBm,
and to Table 3-4 for HP part
numbers.
‘lhble
3-13.
Standard Values for AlOA4R29 and AlOA4R33
T
Resistors
R29 R33
68.1
68.1
68.1 75
75 75
68.1
82.5
75
90.9
3-l 56 Adjustments
Vue de la page 297