2 1. Frequency Response Adjustments
22. On the synthesized sweeper, press
CPOWER
LEVEL] and adjust
the ENTRY knob as necessary for a power meter indication of
-15.00
dBm
*2.00
dB
at 100 MHz.
23. On the power meter, press (RANGE HOLD) (turning it on).
24. On the synthesized sweeper, press (POWER LEVEL) and adjust the
ENTRY knob for a power meter indication of -10.00
dBm
f0.03
dB
at 100 MHz.
25. On the synthesized sweeper, press
[mj
LEVELING and adjust
the ENTRY knob (REF in
dBV
with ATN: 0
dB)
for a power
meter indication of -10.00
dBm
ho.03
dB
at 100 MHz.
Note
Do not vary the synthesized sweeper POWER LEVEL setting
(internal leveling) or METER REF and METER ATN settings (external
power meter leveling) for the remaining steps in this section of the
adjustment procedure. The frequency response adjustments are
referenced to the -10.00
dBm
power level at 100 MHz.
26. Press
[2--22)
on the spectrum analyzer. Set the RF Section
front-panel AMPTD CAL control to the approximate center of its
adjustment range.
Note
Do not vary the spectrum analyzer front-panel AMPTD CAL control
setting for the remaining steps in this adjustment procedure.
27. Set (Band A Lower Segment) A6AllR48 Al and (Band A Upper
Segment) A6AllR66 A2 each to the approximate center of its
adjustment range.
Then, adjust
A6AllR84
GAIN fully clockwise for maximum
RF gain. See Figure 3-90 for the locations of A6AllR48 Al,
A6AllR66 A2, and A6AllR84 GAIN.
28. On the spectrum analyzer, key in
@E5iCiJ
8, [REFERENCE LEVEL)
-4
dBm
and then press the
@)
and
@‘J
keys as necessary to
position the peak of the displayed 100 MHz signal within one
division of the top graticule line. Key in MARKER (PEAK SEARCH),
MARKER
[nl
on the spectrum analyzer to position two markers
on the peak of the displayed 100 MHz signal.
29. Readjust A6AllR84 GAIN counterclockwise to decrease the RF
gain 5.00
dB
*0.02
dB,
as indicated by the MARKER A indication
on the spectrum analyzer display.
30. On
the
spectrum analyzer, key in (RECALL) 8.
31. Adjust
17-turn
potentiometer (Band A Step Gain)
A6AlOR21
GA to adjust the amplitude of the displayed 100 MHz signal to
-
10.00
dBm
*O.
10
dB.
Adjust
A6AlOR21
GA counterclockwise
to increase the signal level, and clockwise to decrease the signal
level. If
A6AlOR21
GA does not have sufficient range, adjust the
amplitude of the 100 MHz displayed signal as close as possible to
-10.00
dBm.
3-190
Adjustments