11. Down/Up Converter Adjustments
23. Connect the tip of the active probe to
A4A4TP7,
and adjust
A4A6AlC31
18.4 MHz NULL for minimum amplitudes of the
displayed 18.4 MHz and 36.8 MHz signals on the second spectrum
analyzer. The level of the displayed 18.4 MHz signal should be
below -10
dBm.
If
A4A6AlC31
has insufficient adjustment range, increase the
value
of factory-select resistor
A4A5RlO.
See Figure 3-49 for the
location of
A4A5RlO.
Refer to
‘Ikble
3-3 for the acceptable range
of values for
A4A5RlO
and to lkble 3-4 for HP part numbers.
Adjustments 3-103