11. Sweep Time Accuracy Test
Procedure
1. Connect equipment as shown in Figure 2-21.
2. Press
[mj
SWEEP
C-1
on the analyzer.
3. Key in the following settings:
[CENTER FREQUENCY_)
.....................................
500 MHz
(FREQUENCY SPAN)
........................................
. . . . .
OHz
4. Set up the universal counter as follows:
a. Set all front panel keys in “out” position.
b. Set POWER switch to ON.
c. Set GATE TIME vernier control to 9 o’clock.
d. Set SEP/COM A switch to COM A position.
e. Depress T.I. A
+
B switch (making sure the blue shift key is
out).
f. Set Channel A trigger level to trigger on negative slope.
g. Set Channel B trigger level to trigger on positive slope.
h. Set both Channel A and Channel B
ac/dc
switches to dc.
i. Connect the digital voltmeter to Channel A TRIGGER LEVEL
OUT. (Be sure to ground the DVM properly.)
j. Adjust Channel A trigger level to set a DVM voltage reading of
0.3 v.
k. Repeat steps i and j for Channel B.
5. Set analyzer @WEEP TIME] to 20 ms. Reset the universal counter
and press SWEEP
C-1
on the spectrum analyzer.
6. Note the measured sweep time on the universal counter and
record this value in
‘Ihble
2-16. The measured sweep time should
be a value between the minimum and maximum values given in
‘Ihble
2-16.
7. Repeat steps 5 and 6 for each sweep time setting in Iable 2-16.
Performance Tests 2-47