
12. Noise Sidebands Test
5.
Key in the following on the analyzer under test:
(CENTER
FREQUENCY]
...................................... 5.7
GHZ
[FREQUENCY SPAN)
..........................................
.5
kHz
6.
7.
8.
Wait for completion of the sweep (the asterisk * at the upper right
of the display will disappear), then press MARKER [PEAK SEARCH),
(j-1,
CMKR
+
REF LVL).
Change [FREQUENCY SPAN) to 700 Hz. Wait for completion of the
sweep, then press MARKER (PEAK SEARCH),
(j-1.
Wait again
for the completion of the sweep.
Press
[SHIFT)
(VIDEoJG
to initiate video averaging. When the
VID AVG readout at the left edge of the display reaches 20, press
m),
TRACE B
(mj.
9.
10.
Press MARKER [PEAK SEARCH], In] and key in 320 Hz.
Record the MARKER A amplitude:
dBc.
See Figure 2-23.
Find the equivalent noise power bandwidth,
BWenp,
for the 10 Hz
resolution bandwidth filter by multiplying the 3
dB
bandwidth
recorded in
Table
2-5 of the Resolution Bandwidth Accuracy Test
by 1.13:
BWen,
=
1.13 x Hz
=
Hz
Figure 2-23. Noise Sidebands Measurement
Performance Tests 2-51
Commentaires sur ces manuels