1. Center Frequency Readout Accuracy Test
Figure 2-2. Center Frequency Accuracy Measurement
Note
The spectrum analyzer CENTER FREQUENCY readout may fall
outside of the specified limits if the internal frequency reference
of the analyzer has not been calibrated within the past year.
To
eliminate the “frequency reference error X center frequency” error,
the analyzer’s 10 MHz Frequency Reference Output (on the rear
panel) may be substituted for the frequency standard.