  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 91
Harmonic
Distortion
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
16. Harmonic and Intermodulation Distortion Test
Set the synthesized sweeper for an output CW frequency of
230.00 MHz and an output level of approximately -30
dBm.
Press
C2J
on the analyzer. Key in the following settings:
CCENTER
FREQUENCY]
.....................................
230 MHz
(FREQUENCY SPAN]
.........................................
100
kHz
(REFERENCE LEVEL) ....................................... -30
dBm
Connect equipment as shown in Figure 2-28, using the 250 MHz
low-pass filter. Adjust the synthesized sweeper output level to
place peak of signal trace at the top graticule line.
Press DISPLAY LINE
(ENTER]
and key in
-
110
dBm.
Press MARKER (PEAK SEARCH],
(j-j,
CMKR/A-+ STP SIZE).
Activate (CENTER FREQUENCY] and press
ch)
to tune to the second
harmonic of the input signal.
Reduce (FREQUENCY SPAN) to 10
kHz
and
c-1
to 30 Hz.
Reduce
(RES]
if necessary, for a margin of 15
dB
between the
displayed noise and the display line.
The second harmonic should be below the display line
(c-80
dBc).
Second harmonic level of 230 MHz
dBc
Replace the 250 MHz low-pass filter with the 1200 MHz low-pass
filter.
Set the synthesized sweeper for an output CW frequency of
800.000 MHz.
Press
(21
on the analyzer. Key in:
[CENTER FREQUENCY)
.....................................
800 MHz
[FREQUENCY
SPAN]
........................................
,100
kHz
[REFERENCE LEVEL)
.......................................
-30
dBm
Adjust synthesized sweeper output level to place peak of signal
trace at the reference level line.
Note
If unable to locate a harmonic distortion product, increase the output
level by 10
dB.
Be sure to return the output level to the original
setting before making a measurement.
13. Press DISPLAY LINE
fjjj
and key in -100
dBm.
14. Press MARKER
CPEAK
SEARCH),
(MKR),
[MKR/A+
STP
SIZE],
[CENTER FREQUENCY)
m,
@EQUENCY
SPAN) 10
kHz.
15. The second harmonic of the input signal should be below the
display line
(c-70
dBc).
Second harmonic level of 800 MHz
dBc
16. Replace the 1200 MHz low-pass filter with the 8
GHz
low-pass
filter.
17. Set the synthesized sweeper for an output frequency of 7200.000
MHz and an output level of 0
dBm.