2. High-Voltage Adjustment (SN 3004A and Above)
AlA
HIGH
VOLTAGE
REGULATOR
AlA8DSl
AlA7TP3
AlA6R103
Figure 3-13. Location of High Voltage Adjustments
4. Set the spectrum analyzer LINE switch to ON. Set the front-panel
INTENSITY control fully counterclockwise (CRT beam at cut-off)
to prevent possible damage to the CRT.
5. Note the DVM indication at AlA7TP3.
DVM Indication:
6. Connect the high-voltage probe to the DVM, and connect the
probe to
AlA7TP3.
7. Note the DVM indication.
DVM Indication:
8. Divide the DVM indication in step 7 by the DVM indication in
step 5. This gives the calibration factor needed to compensate for
high-voltage probe error.
Calibration Factor:
9. Disconnect the high-voltage probe from
AlA7TP3.
Set the
spectrum analyzer LINE switch to STANDBY. Remove the ac line
cord from both instrument sections.
Warning
The MAINS power-on indicator
AlASDSl
(red LED) should
be completely off before proceeding with this procedure. See
Figure 3-13 The indicator will remain lit for several seconds after
the ac line cord has been removed, and will go out slowly (the
light becomes dimmer until it is completely out).
Warning
With the protective cover removed in the following step, do not
place hands near the
AlA
High-Voltage assembly. High voltage
(approximately -2400 V dc) can present even when the ac line
cord is disconnected.
10. Wait at least one minute for capacitors to discharge to a safe
level.
Adjustments 3-43