29.
Set the controls of the second spectrum
anaiyzer
as follows:
CENTER FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . 165 MHz
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.15
MHz
RES BW
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
kHz
REFERENCE LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -7
dBm
LOG SCALE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.I0
dB/DIV
MARKER
. . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . OFF
30.
Press
QNSTR
PRESET) on the synthesized sweeper, and key in
@
330.3 MHz, (POWER LEVEL) 0
dBm.
31.
Use a second BNC to SMB Snap-on Test Cable and an SMB adapter
to connect the cable 4 (yellow) from AlOA3Jl IN 200-300 MHz to
the RF OUTPUT of the synthesized sweeper.
32.
Set the RF Section LINE switch to ON, and key in
L-j,
(CENTER FREQUENCY) 42.57 MHz, (FREQUENCY SPAN) 0 Hz.
17.
20130
Loop Phase Lock Adjustments
Note
Adjustments
AlOA3Lll
165 MHz NULL,
AlOA3L12
160 MHz NULL,
and
AlOA3L13
170 MHz NULL are sealed at the factory and normally
do not require readjustment unless a component failure has occurred.
lb
verify their proper adjustment, skip to step 39.
33.
34.
35.
36.
37.
38.
39.
40.
41.
Adjust
AlOA3Lll
165 MHz NULL,
AlOA3L12
160 MHz NULL, and
AlOA3L13
170 MHz NULL fully clockwise.
Adjust AlOA3Ll3 170 MHz NULL to minimize the amplitude of
the 170 MHz signal displayed on the second spectrum analyzer.
Refer to Figure 3-78 for the location of
AlOA3L3
170 MHz NULL
adjustment.
On the synthesized sweeper, key in Icw) 325.3 MHz.
Adjust
AlOA3Lll
165 MHz NULL to minimize the amplitude of
the 165 MHz signal displayed on the second spectrum analyzer.
Refer to Figure 3-78 for the location of
AlOA3Lll
165 MHz NULL
adjustment.
On the synthesized sweeper, key in Icw) 320.3 MHz.
Adjust
AlOA3L12
160 MHz NULL to minimize the amplitude of
the 160 MHz signal displayed on the second spectrum analyzer.
Refer to Figure 3-78 for the location of
AlOA3L12
160 MHz NULL
adjustment.
On the synthesized sweeper, key in Icw] 260.3 MHz.
On the second spectrum analyzer, press [CENTER FREQUENCY) 100
MHz
CFREQUENCY
SPAN
)
6 MHz MARKER
(PEAK
SEARCH
)
to position a
marker on the peak of the displayed 100 MHz signal, and verify
that the output power level of the
PLLl
IF is at least -14
dBm.
If it is not, a lower
value
can be selected for factory-select
component
AlOA3C26
to improve the impedance match between
the double-balanced mixer
AlOA3Ul
and the IF Input Amplifier
circuit on
AlOA3
PLLl
IF Assembly. Refer to Table 3-3 for the
acceptable range of values for
AlOA3C26,
and to
Table
3-4 for HP
part numbers; refer to Figure 3-78 for the location of
AlOA3C26.
On the synthesized sweeper, key in [START
FREQ]
300.3 MHz.