20.
On the spectrum analyzer, key in the following settings:
(CENTER FREQUENCY)
.....................................
100 MHz
FREQUENCY SPAN)
;&VT)
........................................
.300
kHz
.................................................
.300
kHz
21.
On the spectrum analyzer, press (REFERENCE LEVEL] and use the
DATA knob to position the signal peak near the reference level
(top graticule line).
22.
On the spectrum analyzer, key in (PEAK SEARCH], MARKER [n],
and press
Q)
several times to position the second marker at the
leftmost graticule line.
23.
Adjust
A4A9R62
300
kHz
for a MARKER A indication of -3.00
dB
f0.05
dB.
24.
On the spectrum analyzer, press
ch)
several times to position
the second marker at the rightmost graticule line. Then, press
[STOPFREQJ
and use the DATA knob to adjust the centering of the
displayed signal for a MARKER A indication of -3.00
dB
f0.05
dB.
25.
26.
On the spectrum analyzer, key in (FREQUENCY
SPAN)
300
kHz,
MARKER (OFF), (PEAK SEARCH], MARKER
a],
and press
(JIJ
several
times to position the second marker at the leftmost graticule line.
Readjust
A4A9R62
300
kHz
for a MARKER A indication of -3.00
dB
f0.05
dB.
27.
Repeat steps 24 through 26 as necessary until no further
adjustment is required.
9. 3 dB Bandwidth Adjustments
Note
The 100
kHz
3-dB
bandwidth is set with factory-select components
A4A8R30,
A4A8R55,
A4A4R16,
and
A4A4R60.
If it is necessary to
increase the 100
kHz
3-dB
bandwidth, increase the value of one or
more of these factory-select components. The 30
kHz
3-dB
bandwidth
is set with factory-select components
A4A8R26,
A4A8R52,
A4A4R20,
A4A4R40,
and
A4A4R64.
If it is necessary to increase the 30
kHz
3-dB
bandwidth, decrease the value of one or more of these factory-
select components. Refer to
‘Ihble
3-3 for the acceptable range of
values for
A4A8R30,
A4A8R55,
A4A4R16,
A4A4R60,
A4A8R26,
A4A8R52,
A4A4R20,
A4A4R40,
and
A4A4R64,
and to Table 3-4 for HP
part numbers.
28. On the spectrum analyzer, key in the following settings:
(CENTER FREQUENCY)
.....................................
100 MHZ
FREQUENCY SPAN]
&CiiV,
..........................................
10
kHz
...................................................
10
kHz
29. On the spectrum analyzer, press [REFERENCE LEVEL) and use the
DATA knob to position the signal peak near the reference level
(top graticule line).
30. On the spectrum analyzer, key in (PEAK SEARCH], MARKER (a,
and press
(JJJ
several times to position the second marker at the
leftmost graticule line.
Adjustments 3-91