17.
17. 20/30 Loop Phase Lock Adjustments
Set the RF Section LINE switch to STANDBY. Disconnect the
frequency counter from AlOA5J4 SCAN
I.1
MHz OUT. Reinstall
AlOA6
PLL2
Phase Detector Assembly in the RF Section and
reconnect the cable 7 (violet) to AlOA5J4 SCAN
I.1
MHz OUT.
Set the RF Section LINE switch to ON, and key in
(jZZ@,
(STOP)
10 MHz,
(
START
FREQ)
8.600 MHz, SWEEP
@KKK),
TRACE A (CLEAR-WRITE).
Connect the DVM to
AlOA8TP5
VCO TUNE.
Adjust
AlOA5R2
150 MHz adjustment for a DVM reading of +3.00
f
0.05 Vdc.
Connect the DVM to
AlOA6TP7
PHASE DET OUT, located on the
cover of the
AlOA6
PLL2
Phase Detector Assembly.
Press SWEEP
ICONT).
Adjust
AlOA8R9
.3
MHz adjustment for a
DVM indication of + 3.50
f
0.05 Vdc.
On the RF Section, key in
(
START
FREQ]
8.599 MHz, SWEEP
(GKQ,
TRACE A (CLEAR-WRITE).
Note
The CRT annotation will round off to 8.60 MHz, but
the
RF Section is
actually set to a start frequency of 8.599 MHz.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
Connect the DVM to
AlOA8TP5
VCO TUNE.
Adjust
AlOA5R4
100 MHz adjustment for a DVM indication of
+ 15.00
f
0.05 Vdc.
Connect the DVM to
AlOA6TP7
PHASE DET OUT.
Adjust
AlOA8R41
.2
MHz adjustment for a DVM indication of 3.50
f
0.05 Vdc.
Repeat steps 12 through 21 as necessary until no further
adjustment is required.
On
the
RF Section, connect the front-panel CAL OUTPUT signal to
the RF INPUT.
On the RF Section, key in
(sTopI
100.5 MHz, (START
FREQ]
95.5 MHz, SWEEP
LCONT),
[
SWEEP
TIME)
100 ms.
Adjust
AlOA8R27
5 MHz SCAN to center the 100 MHz CAL
OUTPUT signal on the center graticule line. Refer to Figure 3-79
for the location of
AlOA8R27
5 MHz SCAN adjustment.
On the RF Section, key in
[STOP]
100.05 MHz, (START
FREQ]
99.55 MHz, [SWEEP TIME] 500 msec.
Adjust
AlOA8R25
.5
MHz SCAN to center the 100 MHz CAL
OUTPUT signal on the center graticule line. Refer to Figure 3-79
for the location of
AlOA8R25
.5
MHz SCAN adjustment.
Adjustments
3-l
55