2 1. Frequency Response Adjustments
LOG SCALE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 dB/DIV
187. Press MARKER
(-1
and adjust (IF Step Gain) potentiometer
A6AlOR81
GF for a marker indication of 0.00
dBm
fO.O1
dB.
Note
If
A6AlO
Miscellaneous Bias/Relay Driver Assembly is HP
P/N
85660-60322 (HP 85660AB RF Sections with serial number prefix
2747A or above), and
A6AlOR81
GF does not have sufficient range,
change the
value
of factory-select component
A6AlOR91.
Increase
the
value
of
A6AlOR91
to decrease the signal level, and decrease the
value of
A6AlOR91
to increase the signal level. See
Table
3-3 for the
acceptable range of
values
for
A6AlOR91,
and
‘Ihble
3-4 for HP part
numbers. See Figure 3-90 for the location of
A6AlOR91.
188.
189.
190.
191.
192.
Disconnect the low-loss microwave test cable from the spectrum
analyzer front panel 321.4 MHz IF INPUT connector. Reconnect
the jumper cable between the spectrum analyzer front panel
32 1.4 MHz IF INPUT and IF OUTPUT connectors.
On the spectrum analyzer, key in
m
Q)
u,
6 Hz,
m
(REFERENCE LEVEL)
z
-12
dBm.
In the RF Section, disconnect cable 84 (gray/yellow) from
A6A12Jl
and cable 82 (gray/red) from
A6A12J2.
Replace
the cover to the A6 RF Module, and then reconnect cable 84
(gray/yellow) to
A6A12Jl
and cable 82 (gray/red) to
A6A12J2.
Remove the jumper between
A12TP2
and
A12TP3
(LOCK
INDICATOR DISABLE) on the Al2 Front Panel Interface
Assembly in the RF Section.
Replace the RF Section bottom cover.
Adjustments
3-205