22.
Analog-To-Digital
Converter
Adjustments
Reference
A3A8
Analog-to-Digital Converter
Description
The Analog-to-Digital Ramp Converter is adjusted at zero and
full-scale by injecting a 0 V dc input and + 10 V dc input and adjusting
the OFFS and GAIN controls until the ramp output at A3A8TPll
toggles high to low. This sets the horizontal end points for the CRT
trace display; when the sweep ramp input is at 0 V dc (the left
graticule edge), trace position 1 is set, and when the sweep ramp
input is at + 10 V dc (the right graticule edge), trace position 1000 is
set.
This procedure requires a + 10 V dc source which is stable and
noise-free. A simple supply circuit which can be built with common
components is illustrated in Figure 3-108. If these components are
unavailable, the alternate procedure provided below (using only the
digital voltmeter) can then be used.
LW-NO
I SE
DC SUPPLY
--l
DIGITAL
VOLTWTER
Figure 3-100. Analog-To-Digital Converter Adjustments Setup
Equipment
Oscilloscope
...............................................
HP
545OlA
Digital Voltmeter
...........................................
HP 3456A
Low-Noise DC Supply (Optional)
....................
See Figure 3-108
1O:l
Divider Probe, 10
MG/7.5
pF
........................ .HP
10432A
Procedure
1. Position spectrum analyzer upright as shown in Figure 3-100 and
remove IF-Display Section top cover.
2. Set spectrum analyzer LINE switch to ON and press
C-1.
3. Procedure using Low-Noise DC Supply is illustrated in Figure 3-108.
a.
Key in
(-1
TRACE A and SWEEP
(-1.
b. Disconnect cable 0 (black) from sweep ramp input A3A8Jl.
c. Short
A3A8TP4
to
A3A8TP5
or connect SMB snap-on short to
A3A8Jl.
3-206 Adjustments