1.
Center
Frequency
Readout
Accuracy
‘l&t
Related
Adjustments
10 MHz Standard Adjustment
Sweep, DAC, and Main Coil Driver Adjustments
Specifications
For spans <n X 5 MHz,
f
(2% of frequency span + frequency
reference error X center frequency + 10 Hz).
For spans >n X 5 MHz,
f
(2% of frequency span + n X 100
kHz
+
frequency reference error X center frequency).
n* Center Frequency
1
100 Hz to 5.8
GHz
2 5.8
GHz
to 12.5
GHz
3 12.5
GHz
to 18.6
GHz
4
>18.6
GHz
* n is the harmonic mixing number, depending on center
frequency.
SPECTRUM ANALYZER
FAEOUENCY
STANDARD
Figure 2-l. Center Frequency Test Setup