18.
RF
Module
Phase
Lock
Adjustments
Reference
RF-Section:
A6A9
Phase Lock
Description
The 3.3
GHz
Heterodyne Phase Lock Loop sampler circuits and 300
MHz Tripler circuits on the
A6A9
Phase Lock Assembly are adjusted
and checked for proper operation. The match between the Sampler
Driver circuit and the
A6A9Ul
Sampler is adjusted. Then, the output
balance of the
A6A9Ul
Sampler is set. A second spectrum analyzer is
used to adjust the output match between the Tripler circuit and the
300 MHz Power Amplifier. Tripler match is adjusted for maximum 300
MHz output signal with all harmonics more than 15
dB
down.
Equipment
Spectrum Analyzer
.........................................
HP 8566B
Digitizing Oscilloscope
.................................... HP
545OlA
Digital Voltmeter
...........................................
HP 3456A
15x2 Extender Board (service accessory)
...............
.08505-60041
BNC 50-ohm Termination
.................................
HP
11593A
Cables:
BNC to SMB Cable
(service
accessory) . . . . . . . . . . . .
.85680-60093
Procedure
1. Set the spectrum analyzer LINE switch to STANDBY. Turn the
spectrum analyzer over to position the RF Section on top, as
shown in Figure 3-81, and remove the RF Section bottom cover.
2. In the RF Section, disconnect the cable 84 gray/yellow from
A6A12Jl
and the cable 82 (gray/red) from
A6A12J2.
Remove the
cover from the A6 RF Module, and then reconnect the cable 84
(gray/yellow) to
A6A12Jl
and the cable 82 (gray/red) to
A6A12J2.
Refer to Figure 3-82 for the location of the
A6A12
YTX Driver
Assembly and the A6 RF Module.
Sampler
Match
and
HET
Unlock
Detector
Delay
Adjustments
Caution
A6A9Ul
Sampler is very susceptible to damage from electrostatic
discharge (ESD). Be sure to use proper grounding techniques when
handling
A6A9
Phase Lock assembly and when disconnecting
and connecting cables to
A6A9J5
3.3
GHz
INPUT and test points
A6A9AlE5
and
A6A9AlE6.
3.160