Agilent Technologies HP 8566B Spécifications Page 99

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 405
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 98
17. Image, Multiple, and Out of Rand Responses Test
Procedure
1. Connect equipment as shown in Figure 2-31 with the synthesized
sweeper RF OUTPUT connected to the analyzer input.
2. Press INSTR PRESET on the synthesized sweeper. Key in the
following sweeper settings:
CW FUNCTION
....................................
3000.000 MHz
POWER LEVEL
..........................................
0.0
dBm
3. Press
(2)
on the analyzer. Key in the following analyzer
settings:
[CENTER FREQUENCY)
........................................
3
GHz
[FREQUENCY
SPAN)
.........................................
100
kHz
4. Press DISPLAY LINE
(ENTER)
and key in -70
dBm.
5. Press MARKER
[PRESEL
PEAK) and wait for PEAKING! message to
disappear from the display. Press MARKER
(j).
6. Adjust the power level of the synthesized sweeper to place the
peak of the signal trace at the top CRT graticule line.
Note
If the maximum output power level of the synthesized sweeper is not
enough to place the signal peak at the top CRT graticule line, then
adjust the spectrum analyzer REFERENCE LEVEL
as.required
to
place the signal peak at the top CRT graticule line.
7. Press MARKER
Ia].
Using the DATA knob, determine the
amplitude of the spurious response and enter the result in
‘Ihble
2-20.
8. Set the synthesized sweeper to the frequencies in
‘Pable
2-20
corresponding to an analyzer center frequency of 3
GHz.
The
maximum allowable amplitude of the spurious response at the
analyzer center frequency for each setting is shown in the table.
2-74 Performance Tests
Vue de la page 98
1 2 ... 94 95 96 97 98 99 100 101 102 103 104 ... 404 405

Commentaires sur ces manuels

Pas de commentaire