3. Impulse and Resolution Bandwidth Accuracy Test
Procedure
1. Set the frequency synthesizer for a 15 MHz, + 13
dBm
output.
Connect the output of the frequency synthesizer to the EXT
INPUT of the pulse/function generator.
2. Set the pulse/function generator controls as follows:
MODE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TRIG
EXT INPUT
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . positive-going
EXT INPUT LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . midrange
OUTPUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . pulse
LOL
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ov
HIL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.0.4V
WIDTH (WID) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 ns
DISABLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . off
Note
The spectrum analyzer [REFERENCE LEVEL] setting should remain at
0
dBm
throughout steps 4 through 38 to prevent possible IF gain
compression of the pulse signal.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
On the spectrum analyzer, press
t2-22)
and set the controls as
follows:
.......................................
15 MHz
.......................................
12 MHz
.......................................
..20dB
......................................
3
MHz (i)
..3
MHz
........................................
.......................................
.OdBm
On the spectrum analyzer, press
m,
[ATTEN)
[x)
D,
SWEEP
t-1
MARKER (PEAK SEARCH). Note the MARKER amplitude for
the 3 MHz filter in the HIGH FREQUENCY REPITITION RATE
column in
‘Ihble
4-2.
Set the frequency synthesizer
CFREQUENCY]
to 300
kHz.
On the spectrum analyzer, press [FREQUENCY SPAN] 0 Hz,
[SWEEP TIME] 0.5 seconds, SWEEP
I-1.
Press MARKER
[PEAK
SEARCH
).
Note the MARKER amplitude for
the 3 MHz filter in the LOW FREQUENCY REPITITION RATE
column in
Table
4-2.
Calculate the Impulse Bandwidth of the 3 MHz Alter using the
formula shown below and record the results for the 3 MHz filter
in
‘Ihble
4-2.
BW(i) = High frequency rep rate (15 MHz) x (Low frequency
reading (step
7)/Hi
frequency reading(step 4))
Set the frequency synthesizer (FREQUENCY] to 10 MHz.
On the spectrum analyzer, key in [CENTER FREQUENCY] 10 MHz,
I-1
1 MHz (i),
~FREQUENCY
SPAN
)
4 MHz, SWEEP TIME
(AUTO],
SWEEP
(SINGLE),
MARKER
[
PEAK SEARCH
).
Record MARKER
amplitude in Table 4-2
Set the frequency synthesizer
[FREQUENCY)
to 100
kHz.
On the spectrum analyzer, key in (FREQUENCY
SPAN_)
0 Hz,
[SWEEP TIME) 0.5 seconds, SWEEP