19.
20.
Pulse
Response
of
21.
Control
Gate
Z
Amplifier to
BLANK
Input
22.
23.
Disconnect the oscilloscope channel 4 probe from the spectrum
analyzer. Connect the oscilloscope channel 1 probe to
AlA2TP301,
and connect the probe ground lead to
AlA2TP501.
On the oscilloscope, press
[RECALL)
[m)
to perform a soft reset.
Press
CCHAN],
CHAl&EL
1 on, more preset probe, and use the
front-panel knob to set the probe to
lO.OO:l.
Press more .
24.
Set the oscilloscope controls as follows:
25.
3. Preliminary Display Adjustments (SN 3004A and Above)
Connect the oscilloscope channel 1 probe to
AlA2TP104
and the
channel 4 probe to
AlA2TP105.
See Figure 3-25 for the location
of the test points. Connect the output of the pulse/function
generator to
52
(Y input) on the Display Adjustment PC board in
the
A3A2
slot.
The Y Deflection Amplifier is identical to the X Deflection
Amplifier. Repeat steps 12 through 18 for the Y Deflection
Amplifier using R127, R120, R117,
C104,
and
C109,
respectively.
Press
-1:
amplitude scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.20.0
V/div
offset
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...45.0000 V
Press
I-):
time scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.50.0
ns/div
delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125.000 ns
Press (TRIG]:
level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.50.00000
V
Press
(jjj:
connect dots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
..on
Press
[SHOW).
Connect the output of the Pulse/Function Generator to
53
(Z
input) on the Display Adjustment PC Board in the
A3A2
slot. Set
the board’s switch to the
down
position.
Note
The pulse/function generator output must be terminated with 50
ohms. Use a BNC tee, a
506)
termination, and a BNC female to SMB
female adapter. Install the
500
termination as close to the Display
Adjustment PC Board as possible.
26. Set the pulse/function generator’s controls as follows:
MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . NORM
Waveform
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
..pulse
Frequency (FRQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.200
kHz
Width (WID)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250 ns
Amplitude (AMP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.4.OOV
Offset (OFS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.2.OOV
27. Disconnect the black connector with three wires (8, 98, and 96)
from
AlA2J5,
and set AlA2R319 INT GAIN fully clockwise.
28. Set the spectrum analyzer front-panel INTENSITY control fully
clockwise. Adjust the oscilloscope trigger level for a stable display.
Note the display on the oscilloscope. The pulse should be
L55V
peak-to-peak.
Adjustments 3-61