7. IF Gain Uncertainty Test
2
dB
Gain
Steps
13. Press
[2],
~jG?iTLT)
(FJ.
14. Set [REFERENCE LEVEL] to -1.9
dBm.
15. Press MARKER
(OFFS.
Set
(jVloEoj
to 100 Hz.
16. Set the frequency synthesizer for an output power level of -3.9
dBm.
Set the amplitude increment for 2
dB
steps.
17. Press MARKER
CPEAK
SEARCH),
(KJ
18. Set the analyzer
REFERENCE
LEVEL] and the frequency synthesizer
amplitude according to Table 2-10. At each setting, note the MKR
A amplitude and record it in the table.
‘Ihble
2-10. IF Gain Uncertainty, 2
dB
Steps
(REFERENCE LEVEL) Frequency Deviation
Pm)
Synthesizer (MARKER A
Amplitude
Amplitude
Wm)
(W
-1.9 -3.9
0 (ref)
-3.9 -5.9
-5.9 -7.9
-7.9
-9.9
-9.9 -11.9
L
0.1
dB
Gain
Steps
19. Set [REFERENCE LEVEL) to 0
dB.
20. Set the frequency synthesizer for an output power level of -2.00
dBm.
Set the amplitude increment for 0.1
dB
steps.
21. Press MARKER [PEAK SEARCH),
a.
22. Set the analyzer (REFERENCE
LEVEL]
and the frequency synthesizer
amplitude according to
‘fable
2-11. At each setting, note the MKR
A amplitude and record it in the table.
2-24 Performance Tests