9. Impulse Bandwidth Adjustments
9.
Impulse
Bandwidth
Adjustments
Reference
Related Performance
Test
Description
Equipment
Procedure
IF-Display Section
A4A9
IF Control
Impulse Bandwidth Accuracy Test
The CAL OUTPUT signal is connected to the RF INPUT. Each of the
adjustable resolution bandwidths is selected and adjusted for the
proper impulse bandwidth.
No test equipment is required for this adjustment.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
Position the instrument upright and remove the top cover.
Set the LINE switch to On and press
(-1.
Connect CAL OUTPUT to RF INPUT.
Key in [CENTER FREQUENCY) 100 MHz, [FREQUENCY SPAN] 5 MHz
(RES-
3 MHz, and
m.
Press [REFERENCE LEVEL] and adjust the DATA knob to place the
signal peak near the top CRT graticule. The signal should be
centered about the center line on the graticule.
Press [PEAK SEARCH], MKR
+
(CF),
and MARKER [nl.
Using the DATA knob, adjust the marker down one side of
the display signal to the 7.3
dB
point; CRT MKR A annotation
indicates 0.430 X
Adjust
A4A9R60
3 MHz for MKR la] indication of 1.5 MHz while
maintaining the marker at 0.430 X using the DATA knob. Refer to
Figure 4-8 for the adjustment location.
Press MARKER
Ia].
Adjust the marker to the 7.3
dB
point on the
opposite side of the signal (CRT MKR A annotation indicates 1.00
X. There are now two markers; one on each side of the signal at
the 7.3
dB
point.
CRT MKR A annotation now indicates the impulse bandwidth of
the 3 MHz bandwidth. Impulse bandwidth should be 3.00 MHz
ztO.60
MHz
Key in
(jjj
1 MHz, [FREQUENCY SPAN) 2 MHz,
CPEAK
SEARCH),
and
CMKR).
If necessary, readjust by pressing
CREFERENCE
LEVEL] and using the DATA knob to place the signal
peak near the top of the graticule.
Press MARKER
(OFF)
then MARKER
[n.
Option 462
4-27