10. Frequency Response Test
\
Q
ADAPTER
PDWER
UETER
B
SENSOR
Figure 2-19.
Frequency Response Test Setup (60 MHz to 2.5
GHz,
2 to 22
GEIz)
60
MHz
to 2.5
GHz
36. Connect equipment as shown in Figure 2-19, with one resistive
output of the power splitter connected to the power meter/power
sensor, and the second resistive output connected to the spectrum
analyzer RF INPUT using an APC 3.5 (m) to Type N (m) adapter.
Connect the power meter rear panel RECORDER OUTPUT to the
synthesized sweeper front panel LEVELING EXT INPUT.
37. On the power meter, verify that the [RANGE HOLD) switch is off.
Consult the power sensor Cal Factor versus Frequency graph or
table and set the power meter CAL FACTOR
%
switch to the 100
MHz calibration setting.
38. Press
~NSTR
PRESET] on the synthesized sweeper. Set the controls
of the synthesized sweeper as follows:
CW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 MHz
POWER LEVEL . . . . . . . . . . . . . . . . . . . . . -9.0
dBm
RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
LEVELING . . . . . . . . . . . . . . . . . . . . . . . . . . . . INT
39. On the synthesized sweeper, press (POWER
LEVELS
and adjust the
ENTRY knob for a power meter indication of -15.00
dBm
fO.10
dB
at 100 MHz.
40. On the power meter, press (RANGE HOLD) (turning it on).
41. On the synthesized sweeper, press [POWER LEVEL) and adjust the
ENTRY knob for a power meter indication of -10.00 ho.03
dB
at
100 MHz.
42. On the synthesized sweeper, press
(meter_)
leveling and adjust the
ENTRY knob (REF in
dBV
with ATN: 0
dB)
for a power meter
indication of -10.00
dBm
f0.03
dB
at 100 MHz.
Note
Do not vary the synthesized sweeper POWER LEVEL setting (internal
leveling) or METER REF and METER ATN settings (external power
meter leveling) for the remaining steps in this test procedure.