2 1. Frequency Response Adjustments
verify that the power meter indication is -10.00
dBm
fO.10
dB.
If the 100 MHz CAL OUTPUT power level is not within this
tolerance, perform adjustment procedure 19, “CAL OUTPUT
Adjustment”, before continuing with this adjustment procedure.
6. Disconnect the power meter/power sensor from the
speqtrum
analyzer front-panel CAL OUTPUT connector.
7. Press
(2-22)
on the synthesized sweeper. Set the synthesized
sweeper controls as follows:
CW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.21.4
MHz
POWER LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
-
15.0
dBm
8. Connect the low-loss microwave test cable to the frequency
synthesizer RF OUTPUT using an APC 3.5 (f) to APC 3.5 (f)
adapter. Connect the power meter/power sensor to the opposite
end of the test cable using a Type N (f) to APC 3.5 (f) adapter.
9. On the synthesized sweeper, press (POWER LEVEL) and adjust the
ENTRY knob for a power meter indication of -15.00
dBm
f0.03
dB
at 21.4 MHz.
10. Disconnect cable 81 (gray/brown) Wl from A6A3J3 (21.4 MHz
OUT) on the
A6A3
Last Converter Assembly. Disconnect the
power sensor from the low-loss microwave test cable, and
connect the test cable to cable 81 (gray/brown)
Wl
using an
SMB (m) to SMA (f) adapter. See Figure 3-90 for the location of
A6A3J3 (21.4 MHz OUT).
11. On the spectrum analyzer, key in
[j)
8. Verify that the
displayed signal amplitude indicated by the MARKER is -10.00
dBm
f0.40
dB.
If the displayed signal amplitude is not within
this tolerance, perform the following adjustment procedures
as necessary to adjust the overall RF gain of the HP 85662A
IF/Display Section before continuing with this adjustment
procedure:
5. Log Amplifier Adjustments
6. Video Processor Adjustments
10. Step Gain and 18.4 MHz
Local
Oscillator Adjustments
8. 21.4 MHz Bandwidth Filter Adjustments
11. Down/Up Converter Adjustments
12. Disconnect the low-loss microwave test cable from cable 81
(gray/brown) Wl, and reconnect cable 81 (gray/brown)
Wl
to
A6A3J3 (21.4 MHz OUT)
on
the
A6A3
Last Converter Assembly.
13. Connect the DVM to
AGAlOTPl,
and connect the DVM ground to
A6AlOTP2.
See Figure 3-90 for the location of
AGAlOTPl
and
A6AlOTP2.
14. Adjust
AGAlORl
IO (3.3
GHz
Oscillator Drive, to IE on
A6A4
Second Converter) for a DVM indication of -0.15 ho.01 Vdc. See
Figure 3-90 for the location of
AGAlORl.
15. Connect the DVM to
A6A12TP3
(-9 V), and connect the
DVM ground to
A6A12TP2
(YTX COM) in the RF Section. See
Figure 3-90 for the location of
A6A12TP3
and
A6A12TP2.
3-178 Adjustments