12.
100
MHz
Tripler
13.
Adjustments
14.
15.
16.
17.
18.
19.
20.
18. RF’ Module Phase Lock Adjustments
Disconnect cable 85 (gray/green) from
A6A9J3
100 MHz INPUT,
cable 4 (yellow) from A6A9J4 VCO TUNE, and cable 5 (green)
from A6A9J5 3.3
GHz
INPUT. Remove the
A6A9
Phase Lock
Assembly from the extender board in the RF Section, and replace
the
A6A9
Phase Lock Assembly front cover.
Install the
A6A9
Phase Lock Assembly in the RF Section on
an extender board. Reconnect cable 0 (black) to
A6A9J2
CAL
OUTPUT, cable 85 (gray/green) to A6A9J3 100 MHz INPUT, cable
4 (yellow) to A6A9J4 VCO TUNE, and cable 5 (green) to
A6A9J5
3.3
GHz
INPUT.
Set the spectrum analyzer LINE switch to ON and press
(j-1.
Connect a
50-ohm
termination to the RF Section
front-panel CAL OUTPUT connector.
Press
CsGHz)
on the second spectrum analyzer. Connect the
CAL OUTPUT signal on the second spectrum analyzer to the RF
INPUT on the second spectrum analyzer, and press
[RECALL_)
8.
Adjust the AMPTD CAL control on the second spectrum analyzer
for a -10.00
dBm
displayed signal, and then press
m
[FREQUENCY SPAN)~.
Connect the RF INPUT of the second spectrum analyzer to the RF
Section
A6A9Jl
300 MHz OUTPUT. Set the controls of the second
spectrum analyzer as follows:
CENTER FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.550
MHz
FREQUENCY SPAN
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
GHz
REFERENCE LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
+25
dBm
A’ITEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 dB
LOG dB/DIV
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
dB
The 300 MHz output signal should be visible on the display of
the second spectrum analyzer, along with other harmonics of
100 MHz. Press MARKER
CPEAK
SEARCH
)
to position a marker on
the peak of the displayed 300 MHz signal. Adjust
A6A9AlC29
TRIPLER MATCH to maximize the power level of the 300 MHz
signal.
Press MARKER (PEAK SEARCH) and verify that the power level of
the 300 MHz signal is greater than + 16.5
dBm.
Press MARKER
Ia],
and then press the
m
and
meys
(or turn the DATA knob)
to position the second marker on the peak of each of the other
displayed harmonics in succession, and verify that the level of
each harmonic is greater than 15
dB
down relative to the peak of
the 300 MHz signal.
Disconnect the second spectrum analyzer from the RF Section
A6A9Jl
300 MHz OUTPUT. Remove the
50-ohm
termination from
the RF Section front-panel CAL OUTPUT connector.
Adjustments 3-163