2 1. Frequency Response Adjustments
3.40
dB.
See the TRACE B reference waveform to indicate which
portions of the frequency response waveform require relative
adjustment. 113. On the spectrum analyzer, key in
(ml
4,
m).
113. On the synthesized sweeper, press
Icw)
and enter the frequency
recorded in step 111, positioning the displayed TRACE A signal
response at the highest point on the TRACE B waveform.
Note
The RF Section front-panel AMPTD CAL control should still be set to
the approximate center of its adjustment range from step 26.
114. On the spectrum analyzer, key in MARKER
(j-1,
MARKER
[PEAK SEARCH) to position a marker at the peak of the displayed
TRACE A signal response. Adjust 17-turn (Band C Step Gain)
potentiometer
A6AlOR25
GC to adjust the-amplitude of the
marker to -10.00
dBm
+ (l/2 of Band C total peak-to-peak
deviation)
f0.01
dB.
Adjust
A6AlOR25
GC counterclockwise to
increase the signal level, and clockwise to decrease the signal
level.
Note
If
A6AlO
Miscellaneous Bias/Relay Driver Assembly is HP
P/N
85660-60322 (HP 85660AB RF Sections with serial number prefix
2747A or above), and
A6AlOR25
GC does not have sufficient range,
change the
value
of factory-select component
A6AlOR88.
Increase
the
value
of
A6AlOR88
to decrease the signal level, and decrease the
value of
A6AlOR88
to increase the signal level. See
‘Ihble
3-3 for the
acceptable range of values for
A6AlOR88,
and
‘Iable
3-4 for HP part
numbers. See Figure 3-90 for the location of
A6AlOR88.
Band
D,
12.5
to
18.6
GHz
115. On the spectrum analyzer, key in
[WI,
(SHIFT)
(PRESEL
PEAK)
=
, (SHIFT) TRACE A
(jhnAx)
b.
Set the spectrum analyzer
controls as follows:
START FREQ
..........................................
.12.5
GHz
STOP FREQ
............................................
18.6
GHz
RES BW
..................................................
.3
MHz
REFERENCE LEVEL
...................................
-7
dBm
LOG SCALE
..........................................
10 dB/DIV
SWEEP TIME
............................................
500 ms
116. On the synthesized sweeper, key in
@
15.0
GHz.
Note
The POWER LEVEL of the synthesized sweeper should still be leveled
to a calibrated -10.00
dBm
at 100 MHz at the spectrum analyzer RF
INPUT from step 55.
117.
118.
Adjust 22-turn (IF Offset) potentiometer
A6A12R83
D and
17-turn (YTX Bias) potentiometer
A6AlOR12
VD for maximum
signal amplitude at 15.0
GHz
on the spectrum analyzer display.
See Figure 3-90 for the locations of
A6A12R83
D and
A6AlOR15
VD.
On the spectrum analyzer, key in MARKER (NORMAL),
MARKER
[PEAKSEARCH)
to position a marker at the peak of
3-194 Adjustments
Vue de la page 333